Technical Aspects and Validation of a New Biofeedback System for Measuring Lower Limb Loading in the Dynamic Situation

Marco Raaben*, Herman R. Holtslag, Robin Augustine, Rutger O. van Merkerk, Bart F. J. M. Koopman, Taco J. Blokhuis

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

Background: A variety of techniques for measuring lower limb loading exists, each with their own limitations. A new ambulatory biofeedback system was developed to overcome these limitations. In this study, we described the technical aspects and validated the accuracy of this system. Methods: A bench press was used to validate the system in the static situation. Ten healthy volunteers were measured by the new biofeedback system and a dual-belt instrumented treadmill to validate the system in the dynamic situation. Results: Bench press results showed that the sensor accurately measured peak loads up to 1000 N in the static situation. In the healthy volunteers, the load curves measured by the biofeedback system were similar to the treadmill. However, the peak loads and loading rates were lower in the biofeedback system in all participants at all speeds. Conclusions: Advanced sensor technologies used in the new biofeedback system resulted in highly accurate measurements in the static situation. The position of the sensor and the design of the biofeedback system should be optimized to improve results in the dynamic situation.

Original languageEnglish
Article number658
Number of pages10
JournalSensors
Volume17
Issue number3
DOIs
Publication statusPublished - Mar 2017

Keywords

  • biofeedback
  • ambulatory monitoring
  • lower limb loading
  • weight-bearing
  • PRESSURE MEASUREMENT SYSTEM
  • WEIGHT-BEARING
  • TREADMILL WALKING
  • SENSOR SYSTEM

Cite this