TY - JOUR
T1 - Prognostic Value of Thrombus Volume and Interaction With First-Line Endovascular Treatment Device Choice
AU - van Voorst, Henk
AU - Bruggeman, Agnetha A. E.
AU - Andriessen, Jurr W.
AU - Hoving, Jan R.
AU - Konduri, Praneeta
AU - Yang, Wenjin
AU - Kappelhof, Manon
AU - Terreros, Nerea Arrarte
AU - Roos, Yvo B. W. E. M. H.
AU - van Zwam, Wim
AU - van der Lugt, Aad
AU - van der Hoorn, Anouk
AU - Boiten, Jelis
AU - Roosendaal, Stefan
AU - Jenniskens, Sjoerd
AU - Caan, Matthan W. A. A.
AU - Marquering, Henk J.
AU - Emmer, Bart
AU - Majoie, Charles B. L. M.
PY - 2023/4/1
Y1 - 2023/4/1
N2 - Background:A larger thrombus in patients with acute ischemic stroke might result in more complex endovascular treatment procedures, resulting in poorer patient outcomes. Current evidence on thrombus volume and length related to procedural and functional outcomes remains contradicting. This study aimed to assess the prognostic value of thrombus volume and thrombus length and whether this relationship differs between first-line stent retrievers and aspiration devices for endovascular treatment. Methods:In this multicenter retrospective cohort study, 670 of 3279 patients from the MR CLEAN Registry (Multicenter Randomized Clinical Trial of Endovascular Treatment for Acute Ischemic Stroke in the Netherlands) for endovascularly treated large vessel occlusions were included. Thrombus volume (0.1 mL) and length (0.1 mm) based on manual segmentations and measurements were related to reperfusion grade (expanded Treatment in Cerebral Infarction score) after endovascular treatment, the number of retrieval attempts, symptomatic intracranial hemorrhage, and a shift for functional outcome at 90 days measured with the reverted ordinal modified Rankin Scale (odds ratio >1 implies a favorable outcome). Univariable and multivariable linear and logistic regression were used to report common odds ratios (cORs)/adjusted cOR and regression coefficients (B/aB) with 95% CIs. Furthermore, a multiplicative interaction term was used to analyze the relationship between first-line device choice, stent retrievers versus aspiration device, thrombus volume, and outcomes. Results:Thrombus volume was associated with functional outcome (adjusted cOR, 0.83 [95% CI, 0.71-0.97]) and number of retrieval attempts (aB, 0.16 [95% CI, 0.16-0.28]) but not with the other outcome measures. Thrombus length was only associated with functional independence (adjusted cOR, 0.45 [95% CI, 0.24-0.85]). Patients with more voluminous thrombi had worse functional outcomes if endovascular treatment was based on first-line stent retrievers (interaction cOR, 0.67 [95% CI, 0.50-0.89]; P=0.005; adjusted cOR, 0.74 [95% CI, 0.55-1.0]; P=0.04). Conclusions:In this study, patients with a more voluminous thrombus required more endovascular thrombus retrieval attempts and had a worse functional outcome. Patients with a lengthier thrombus were less likely to achieve functional independence at 90 days. For more voluminous thrombi, first-line stent retrieval compared with first-line aspiration might be associated with worse functional outcome.
AB - Background:A larger thrombus in patients with acute ischemic stroke might result in more complex endovascular treatment procedures, resulting in poorer patient outcomes. Current evidence on thrombus volume and length related to procedural and functional outcomes remains contradicting. This study aimed to assess the prognostic value of thrombus volume and thrombus length and whether this relationship differs between first-line stent retrievers and aspiration devices for endovascular treatment. Methods:In this multicenter retrospective cohort study, 670 of 3279 patients from the MR CLEAN Registry (Multicenter Randomized Clinical Trial of Endovascular Treatment for Acute Ischemic Stroke in the Netherlands) for endovascularly treated large vessel occlusions were included. Thrombus volume (0.1 mL) and length (0.1 mm) based on manual segmentations and measurements were related to reperfusion grade (expanded Treatment in Cerebral Infarction score) after endovascular treatment, the number of retrieval attempts, symptomatic intracranial hemorrhage, and a shift for functional outcome at 90 days measured with the reverted ordinal modified Rankin Scale (odds ratio >1 implies a favorable outcome). Univariable and multivariable linear and logistic regression were used to report common odds ratios (cORs)/adjusted cOR and regression coefficients (B/aB) with 95% CIs. Furthermore, a multiplicative interaction term was used to analyze the relationship between first-line device choice, stent retrievers versus aspiration device, thrombus volume, and outcomes. Results:Thrombus volume was associated with functional outcome (adjusted cOR, 0.83 [95% CI, 0.71-0.97]) and number of retrieval attempts (aB, 0.16 [95% CI, 0.16-0.28]) but not with the other outcome measures. Thrombus length was only associated with functional independence (adjusted cOR, 0.45 [95% CI, 0.24-0.85]). Patients with more voluminous thrombi had worse functional outcomes if endovascular treatment was based on first-line stent retrievers (interaction cOR, 0.67 [95% CI, 0.50-0.89]; P=0.005; adjusted cOR, 0.74 [95% CI, 0.55-1.0]; P=0.04). Conclusions:In this study, patients with a more voluminous thrombus required more endovascular thrombus retrieval attempts and had a worse functional outcome. Patients with a lengthier thrombus were less likely to achieve functional independence at 90 days. For more voluminous thrombi, first-line stent retrieval compared with first-line aspiration might be associated with worse functional outcome.
KW - functional status
KW - ischemic stroke
KW - prognosis
KW - stents
KW - thrombosis
KW - THROMBECTOMY
KW - STROKE
KW - RECANALIZATION
KW - LENGTH
KW - GUIDELINES
KW - IMPACT
KW - HEART
U2 - 10.1161/STROKEAHA.122.041606
DO - 10.1161/STROKEAHA.122.041606
M3 - Article
C2 - 36912141
SN - 0039-2499
VL - 54
SP - 1056
EP - 1065
JO - Stroke
JF - Stroke
IS - 4