Abstract
We discuss ion extraction from an internal target in a storage ring, separation of atomic and molecular species, and subsequent measurement of the polarization of the atomic ions. This technique can be applied to internal targets of polarized hydrogen, deuterium, He-3 and other gases. Here, we report on the development of such a device for a tensor polarized deuterium gas target internal to a medium-energy electron storage ring. Ions were extracted from a storage cell using a spherical deflector and transported to a mass analyzer. In a prototype setup, using electrons from a gun to simulate the electron beam from a storage ring, the tensor polarization of the atomic ions was measured using the H-3((d) over right arrow,n)alpha reaction. The data show that the extraction of ions from a storage-cell target, followed by their mass and polarization analysis, allows the determination of the nuclear polarization of the target independent of its spatial and temporal variations.
Original language | English |
---|---|
Pages (from-to) | 212-220 |
Number of pages | 9 |
Journal | Nuclear Instruments & Methods in Physics Research Section A-Accelerators Spectrometers Detectors and Associated Equipment |
Volume | 379 |
Issue number | 2 |
DOIs | |
Publication status | Published - 11 Sept 1996 |
Externally published | Yes |
Keywords
- GAS TARGETS
- STORAGE CELL
- HYDROGEN
- DEUTERIUM
- FIELD
- BEAMS
- RING