Evaluation of a cone-beam computed tomography system calibrated for accurate radiotherapy dose calculation

Marta Bogowicz, Didier Lustermans*, Vicki Trier Taasti, Colien Hazelaar, Frank Verhaegen, Gabriel Paiva Fonseca, Wouter van Elmpt

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

Background and purpose: Dose calculation on cone-beam computed tomography (CBCT) images has been less accurate than on computed tomography (CT) images due to lower image quality and discrepancies in CT numbers for CBCT. As increasing interest arises in offline and online re-planning, dose calculation accuracy was evaluated for a novel CBCT imager integrated into a ring gantry treatment machine. Materials and methods: The new CBCT system allowed fast image acquisition (5.9 s) by using new hardware, including a large-size flat panel detector, and incorporated image-processing algorithms with iterative reconstruction techniques, leading to accurate CT numbers allowing dose calculation. In this study, CBCT- and CT-based dose calculations were compared based on three anthropomorphic phantoms, after CBCT-to-mass-density calibration was performed. Six plans were created on the CT scans covering various target locations and complexities, followed by CBCT to CT registrations, copying of contours, and re-calculation of the plans on the CBCT scans. Dose-volume histogram metrics for target volumes and organs-at-risk (OARs) were evaluated, and global gamma analyses were performed. Results: Target coverage differences were consistently below 1.2 %, demonstrating the agreement between CT and re-calculated CBCT dose distributions. Differences in Dmean for OARs were below 0.5 Gy for all plans, except for three OARs, which were below 0.8 Gy (<1.1 %). All plans had a 3 %/1mm gamma pass rate > 97 %. Conclusions: This study demonstrated comparable results between dose calculations performed on CBCT and CT acquisitions. The new CBCT system with enhanced image quality and CT number accuracy opens possibilities for off-line and on-line re-planning.
Original languageEnglish
Article number100566
Number of pages7
JournalPhysics & Imaging in Radiation Oncology
Volume29
DOIs
Publication statusPublished - Jan 2024

Keywords

  • Cone-beam computed tomography
  • CT number accuracy
  • Dose calculation
  • Image quality
  • Metal artifact reduction
  • Radiotherapy