Accuracy and precision of CPET equipment: A comparison of breath-by-breath and mixing chamber systems.

C. Beijst, G. Schep, E. van Breda, P.F. Wijn, C.V. Pul

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

Cardiopulmonary exercise testing (CPET) has become an important diagnostic tool for patients with cardiorespiratory disease and can monitor athletic performance measuring maximal oxygen uptake [Formula: see text]Vo2(; max). The aim of this study is to compare the accuracy and precision of a breath-by-breath and a mixing chamber CPET system, using two methods. First, this study developed a (theoretical) error analysis based on general error propagation theory. Second, calibration measurements using a metabolic simulator were performed. Error analysis shows that the error in oxygen uptake ([Formula: see text]Vo2) and carbon dioxide production (Vco2[Formula: see text]) is smaller for mixing chamber than for breath-by-breath systems. In general, the error of the flow sensor [Formula: see text]deltaV, the error in temperature of expired air deltaT(B) and the delay time error deltat(delay) are significant sources of error. Measurements using a metabolic simulator show that breath-by-breath systems are less stabile for different values of minute ventilation than mixing chamber systems.
Original languageEnglish
Pages (from-to)35-42
JournalJournal of Medical Engineering & Technology
Volume37
Issue number1
DOIs
Publication statusPublished - 1 Jan 2013