Radiostereometry (RSA) of the cemented Scientific Hip Prosthesis (SHP) reported excessive migration and predicted high failure rates. In a prospective randomized clinical trial we compared minimum 10 years results of the SHP (n=38) with the Omnifit-stem (n=37). Two-year bone remodelling, compared with dual energy x-ray absorptiometry and assessed in regions of interest A-D based on the 7 Gruen zones, showed better periprosthetic bone preservation around the SHP in all but one regions (P