Active Breathing Control in Combination With Ultrasound Imaging: A Feasibility Study of Image Guidance in Stereotactic Body Radiation Therapy of Liver Lesions

Esther Bloemen-van Gurp, Skadi van der Meer, Janet Hendry, Jeroen Buijsen, Peter Visser, Davide Fontanarosa, Martin Lachaine, Guido Lammering, Frank Verhaegen*

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

Purpose: Accurate tumor positioning in stereotactic body radiation therapy (SBRT) of liver lesions is often hampered by motion and setup errors. We combined 3-dimensional ultrasound imaging (3DUS) and active breathing control (ABC) as an image guidance tool. Methods and Materials: We tested 3DUS image guidance in the SBRT treatment of liver lesions for 11 patients with 88 treatment fractions. In 5 patients, 3DUS imaging was combined with ABC. The uncertainties of US scanning and US image segmentation in liver lesions were determined with and without ABC. Results: In free breathing, the intraobserver variations were 1.4 mm in left-right (L-R), 1.6 mm in superior-inferior (S-I), and 1.3 mm anterior-posterior (A-P). and the interobserver variations were 1.6 mm (L-R), 2.8 mm (S-I), and 1.2 mm (A-P). The combined uncertainty of US scanning and matching (inter-and intraobserver) was 4 mm (1 SD). The combined uncertainty when ABC was used reduced by 1.7 mm in the S-I direction. For the L-R and A-P directions, no significant difference was observed. Conclusion: 3DUS imaging for IGRT of liver lesions is feasible, although using anatomic surrogates in the close vicinity of the lesion may be needed. ABC-based breath-hold in midventilation during 3DUS imaging can reduce the uncertainty of US-based 3D table shift correction.
Original languageEnglish
Pages (from-to)1096-1102
JournalInternational Journal of Radiation Oncology Biology Physics
Volume85
Issue number4
DOIs
Publication statusPublished - 15 Mar 2013

Cite this