Time dependent pre-treatment EPID dosimetry for standard and FFF VMAT

Mark Podesta*, Sebastiaan M. J. J. G. Nijsten, Lucas C. G. G. Persoon, Stefan G. Scheib, Christof Baltes, Frank Verhaegen

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

Methods to calibrate Megavoltage electronic portal imaging devices (EPIDs) for dosimetry have been previously documented for dynamic treatments such as intensity modulated radiotherapy (IMRT) using flattened beams and typically using integrated fields. While these methods verify the accumulated field shape and dose, the dose rate and differential fields remain unverified. The aim of this work is to provide an accurate calibration model for time dependent pre-treatment dose verification using amorphous silicon (a-Si) EPIDs in volumetric modulated arc therapy (VMAT) for both flattened and flattening filter free (FFF) beams. A general calibration model was created using a Varian TrueBeam accelerator, equipped with an aS1000 EPID, for each photon spectrum 6 MV, 10 MV, 6 MV-FFF, 10 MV-FFF. As planned VMAT treatments use control points (CPs) for optimization, measured images are separated into corresponding time intervals for direct comparison with predictions. The accuracy of the calibration model was determined for a range of treatment conditions. Measured and predicted CP dose images were compared using a time dependent gamma evaluation using criteria (3%, 3 mm, 0.5 sec). Time dependent pre-treatment dose verification is possible without an additional measurement device or phantom, using the on-board EPID. Sufficient data is present in trajectory log files and EPID frame headers to reliably synchronize and resample portal images. For the VMAT plans tested, significantly more deviation is observed when analysed in a time dependent manner for FFF and non-FFF plans than when analysed using only the integrated field. We show EPID-based pre-treatment dose verification can be performed on a CP basis for VMAT plans. This model can measure pre-treatment doses for both flattened and unflattened beams in a time dependent manner which highlights deviations that are missed in integrated field verifications.
Original languageEnglish
Pages (from-to)4749-4768
JournalPhysics in Medicine and Biology
Volume59
Issue number16
DOIs
Publication statusPublished - 21 Aug 2014

Keywords

  • VMAT
  • time dependent
  • EPID dosimetry
  • control point

Fingerprint

Dive into the research topics of 'Time dependent pre-treatment EPID dosimetry for standard and FFF VMAT'. Together they form a unique fingerprint.

Cite this