Replacement Adaptor 09106 for patients with a dynamic graciloplasty or patients with sacral neuromodulation and abdominal implantable pulse generators: a retrospective, single centre, Stage 2a/2b development IDEAL case series

Roman Assmann*, Perla Douven, Elbert A Joosten, Gommert A van Koeveringe, Stephanie O Breukink, Jarno Melenhorst

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

AIM: Due to the introduction of a new implantable pulse generator (IPG), the Interstim II, patients with either a dynamic graciloplasty or an abdominally placed IPG for sacral neuromodulation could not undergo surgery to replace their IPG in case of end of battery life. For these patients, the Medtronic Replacement Adaptor 09106 was created. This retrospective case series aims to study safety and feasibility of the Medtronic Replacement Adaptor 09106 in patients with abdominally placed IPGs.

METHODS: Seventeen patients (11 female, 6 male) received a replacement adaptor with a follow-up of 6 months. Outcome measures consisted of a bowel habit diary. Adverse events were classified using the Clavien-Dindo classification.

RESULTS: Outcome measures in the bowel habit diaries after replacement (feasibility) did not differ significantly from outcome measures before replacement. Adverse events occurred in 4 out of 17 patients (24%): 2 patients initially showed pocket site pain (Clavien-Dindo level I), which resolved without intervention. One patient suffered from poor wound closure (Clavien-Dindo level II), and 1 patient had persisting pocket pain (Clavien-Dindo level IIIa), for which a pocket revision was performed. Statistical analyses were performed making paired comparisons using a Wilcoxon signed rank test.

CONCLUSION: The Medtronic Replacement Adaptor 09106 is a valuable option for patients with dynamic graciloplasty or sacral neuromodulation and abdominal IPG and has complication rates similar to replacement of the Interstim without Replacement Adaptor 09106.

Original languageEnglish
Pages (from-to)150-155
Number of pages6
JournalColorectal Disease
Volume25
Issue number1
Early online date13 Oct 2022
DOIs
Publication statusPublished - 2023

Fingerprint

Dive into the research topics of 'Replacement Adaptor 09106 for patients with a dynamic graciloplasty or patients with sacral neuromodulation and abdominal implantable pulse generators: a retrospective, single centre, Stage 2a/2b development IDEAL case series'. Together they form a unique fingerprint.

Cite this