T1 - Point-of-Care Test for Detection of Urogenital Chlamydia in Women Shows Low Sensitivity. A Performance Evaluation Study in Two Clinics in Suriname
AU - van der Helm, Jannie J.
AU - Sabajo, Leslie O. A.
AU - Grunberg, Antoon W.
AU - Morre, Servaas A.
AU - Speksnijder, Arjen G. C. L.
AU - de Vries, Henry J. C.
PY - 2012/2/29
Y1 - 2012/2/29
N2 - Background: In general, point-of-care (POC) tests for Chlamydia trachomatis (Ct) show disappointing test performance, especially disappointing sensitivity results. However, one study sponsored by the manufacturer (Diagnostics for the Real World) reported over 80% sensitivity with their Chlamydia Rapid Test (CRT). We evaluated the performance of this CRT in a non-manufacturer-sponsored trial. Methods: Between July 2009 and February 2010, we included samples from 912 women in both high-and low-risk clinics for sexually transmitted infections (STIs) in Paramaribo, Suriname. Sensitivity, specificity, positive-and negative predictive values (PPV and NPV) for CRT compared to NAAT (Aptima, Gen-Probe) were determined. Quantitative Ct load and human cell load were determined in all CRT and/or NAAT positive samples. Results: CRT compared to NAAT showed a sensitivity and specificity of 41.2% (95% CI, 31.9%-50.9%) and 96.4% (95% CI, 95.0%-97.5%), respectively. PPV and NPV were 59.2% (95% CI, 47.5%-70.1%) and 92.9% (95% CI, 91.0%-94.5%), respectively. Quantitative Ct bacterial load was 73 times higher in NAAT-positive/CRT-positive samples compared to NAAT-positive/CRT-negative samples (p <0.001). Human cell load did not differ between true-positive and false-negative CRT results (p = 0.835). Sensitivity of CRT in samples with low Ct load was 12.5% (95% CI, 5.2%-24.2%) and in samples with high Ct load 73.5% (95% CI, 59.9%-84.4%). Conclusions: The sensitivity of CRT for detecting urogenital Ct in this non-manufacturer-sponsored study did not meet the expectations as described previously. The CRT missed samples with a low Ct load. Improved POC are needed as meaningful diagnostic to reduce the disease burden of Ct.
AB - Background: In general, point-of-care (POC) tests for Chlamydia trachomatis (Ct) show disappointing test performance, especially disappointing sensitivity results. However, one study sponsored by the manufacturer (Diagnostics for the Real World) reported over 80% sensitivity with their Chlamydia Rapid Test (CRT). We evaluated the performance of this CRT in a non-manufacturer-sponsored trial. Methods: Between July 2009 and February 2010, we included samples from 912 women in both high-and low-risk clinics for sexually transmitted infections (STIs) in Paramaribo, Suriname. Sensitivity, specificity, positive-and negative predictive values (PPV and NPV) for CRT compared to NAAT (Aptima, Gen-Probe) were determined. Quantitative Ct load and human cell load were determined in all CRT and/or NAAT positive samples. Results: CRT compared to NAAT showed a sensitivity and specificity of 41.2% (95% CI, 31.9%-50.9%) and 96.4% (95% CI, 95.0%-97.5%), respectively. PPV and NPV were 59.2% (95% CI, 47.5%-70.1%) and 92.9% (95% CI, 91.0%-94.5%), respectively. Quantitative Ct bacterial load was 73 times higher in NAAT-positive/CRT-positive samples compared to NAAT-positive/CRT-negative samples (p <0.001). Human cell load did not differ between true-positive and false-negative CRT results (p = 0.835). Sensitivity of CRT in samples with low Ct load was 12.5% (95% CI, 5.2%-24.2%) and in samples with high Ct load 73.5% (95% CI, 59.9%-84.4%). Conclusions: The sensitivity of CRT for detecting urogenital Ct in this non-manufacturer-sponsored study did not meet the expectations as described previously. The CRT missed samples with a low Ct load. Improved POC are needed as meaningful diagnostic to reduce the disease burden of Ct.
U2 - 10.1371/journal.pone.0032122
DO - 10.1371/journal.pone.0032122
M3 - Article
C2 - 22393383
SN - 1932-6203
VL - 7
JO - PLOS ONE
JF - PLOS ONE
IS - 2
M1 - e32122
ER -