Abstract
Objective To construct a mediastinal-specific fluorine-18-fluorodeoxyglucose (F-18-FDG)-PET/MR protocol with high-quality MRI of minimal acquisition-time and comparable diagnostic value to F-18-FDG-PET/computed tomography (CT).
Materials and methods Fifteen healthy participants received PET/MRI and 10 patients with mediastinal tumours (eight non-small-cell lung, two oesophageal cancer) received F-18-FDG-PET/MRI immediately after F-18-FDG-PET/CT. Sequences volume interpolated breath-hold examination (T1-VIBE) and Half-Fourier acquisition single-shot turbo spin echo (T2-HASTE) were optimised by varying the parameters: breath-hold (BH, end-expiration), fat suppression (spectral adiabatic inversion recovery), and ECG-triggering (ECG, end-diastole). Image quality (IQ) of each sequence-variation was qualitatively scored by medical experts and quantitatively assessed by calculating signal-to-noise ratios, contrast relative to muscle, standardized-uptake-value, and tumour-to-blood ratios. Patient comfort was evaluated on patients' experience. Diagnostic accuracy of F-18-FDG-PET/MRI was compared to F-18-FDG-PET/CT, in reference to histopathology/cytopathology.
Results ECG-triggered T1-VIBE images showed the highest signal-to-noise ratio (P <0.01) and the largest contrast between mediastinal soft-tissues, regardless of BH or free-breathing acquisition. IQ of ECG-triggered T1-VIBE scans in BH were scored qualitatively highest with good reader agreement (kappa = 0.62). IQ of T2-HASTE was not significantly affected by BH acquisition (P > 0.9). Qualitative IQ of T1-VIBE and T2-HASTE declined after spectral adiabatic inversion recovery fat-suppression. All patients could maintain BH at end-expiration and reported no discomfort. Diagnostic performance of F-18-FDG-PET/MR was not significantly different from F-18-FDG-PET/CT with comparable staging, standardized-uptake-values, and tumour-to-blood ratios. However, T-status was more often over-staged on F-18-FDG-PET/CT, while N-status was more frequently under-staged on F-18-FDG-PET/MR.
Conclusion ECG-triggered T1-VIBE sequences acquired during short, multiple BHs are recommended for mediastinal imaging using F-18-FDG-PET/MR. With dedicated protocols, F-18-FDG-PET/MRI will be useful in thoracic oncology and aid in diagnostic evaluation and tailored treatment decision-making. Copyright (C) 2019 Wolters Kluwer Health, Inc. All rights reserved.
Original language | English |
---|---|
Pages (from-to) | 815-826 |
Number of pages | 12 |
Journal | Nuclear Medicine Communications |
Volume | 40 |
Issue number | 8 |
DOIs | |
Publication status | Published - Aug 2019 |
Keywords
- hybrid imaging
- mediastinum
- non-small-cell lung cancer
- oesophageal cancer
- PET-CT
- PET-MRI
- CELL LUNG-CANCER
- ESOPHAGEAL CANCER
- INTERNATIONAL WORKSHOP
- PRACTICE GUIDELINES
- COMBINED PET/MRI
- ECHO-SEQUENCE
- RADIAL VIBE
- PET/CT
- MRI
- MOTION