Performance of the electromagnetic calorimeter of the HERMES experiment

H. Avakian, N. Bianchi, G.P. Capitani, E. De Sanctis, P. Di Nezza, A. Fantoni*, V. Giourdjian, R. Mozzetti, V. Muccifora, M. Nupieri, A.R. Reolon, P. Rossi, J.F.J. van den Brand, M. Doets, T. Henkes, M. Kolstein, A. Airapetian, N. Akopov, M. Amarian, R. AvakianA. Avetissian, V. Garibian, S. Taroian

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

The performance of the electromagnetic calorimeter of the HERMES experiment is described. The calorimeter consists of 840 radiation resistant F101 lead-glass counters. The response to positrons up to 27.5 GeV, the comparison between the measured energy and the momentum reconstructed from tracking, long-term stability, hadron rejection and neutral meson invariant mass reconstruction are shown. (C) 1998 Published by Elsevier Science B.V. All rights reserved.