Left Ventricular Myocardial Septal Pacing in Close Proximity to LBB Does Not Prolong the Duration of the Left Ventricular Lateral Wall Depolarization Compared to LBB Pacing

Karol Curila*, Pavel Jurak, Kevin Vernooy, Marek Jastrzebski, Petr Waldauf, Frits Prinzen, Josef Halamek, Marketa Susankova, Lucie Znojilova, Radovan Smisek, Jakub Karch, Filip Plesinger, Pawel Moskal, Luuk Heckman, Jan Mizner, Ivo Viscor, Vlastimil Vondra, Pavel Leinveber, Pavel Osmancik

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

Background: Three different ventricular capture types are observed during left bundle branch pacing (LBBp). They are selective LBB pacing (sLBBp), non-selective LBB pacing (nsLBBp), and myocardial left septal pacing transiting from nsLBBp while decreasing the pacing output (LVSP). Study aimed to compare differences in ventricular depolarization between these captures using ultra-high-frequency electrocardiography (UHF-ECG). Methods: Using decremental pacing voltage output, we identified and studied nsLBBp, sLBBp, and LVSP in patients with bradycardia. Timing of ventricular activations in precordial leads was displayed using UHF-ECGs, and electrical dyssynchrony (e-DYS) was calculated as the difference between the first and last activation. The durations of local depolarizations (Vd) were determined as the width of the UHF-QRS complex at 50% of its amplitude. Results: In 57 consecutive patients, data were collected during nsLBBp (n = 57), LVSP (n = 34), and sLBBp (n = 23). Interventricular dyssynchrony (e-DYS) was significantly lower during LVSP -16 ms (-21; -11), than nsLBBp -24 ms (-28; -20) and sLBBp -31 ms (-36; -25). LVSP had the same V1d-V8d as nsLBBp and sLBBp except for V3d, which during LVSP was shorter than sLBBp; the mean difference -9 ms (-16; -1), p = 0.01. LVSP caused less interventricular dyssynchrony and the same or better local depolarization durations than nsLBBp and sLBBp irrespective of QRS morphology during spontaneous rhythm or paced QRS axis. Conclusions: In patients with bradycardia, LVSP in close proximity to LBB resulted in better interventricular synchrony than nsLBBp and sLBBp and did not significantly prolong depolarization of the left ventricular lateral wall.

Original languageEnglish
Article number787414
Number of pages10
JournalFrontiers in Cardiovascular Medicine
Volume8
DOIs
Publication statusPublished - 7 Dec 2021

Keywords

  • left bundle branch pacing
  • left septal myocardial pacing
  • UHF-ECG
  • dyssynchrony
  • depolarization duration