Abstract
PURPOSE: In order to fully benefit from the improved signal-to-noise and contrast-to-noise ratios at 9.4T, the challenges of B1+ inhomogeneity and the long acquisition time of high-resolution 2D gradient-recalled echo (GRE) imaging were addressed.
THEORY AND METHODS: Flip angle homogenized excitations were achieved by parallel transmission (pTx) of 3-spoke pulses, designed by magnitude least-squares optimization in a slice-by-slice fashion; the acquisition time reduction was achieved by simultaneous multislice (SMS) pulses. The slice-specific spokes complex radiofrequency scaling factors were applied to sinc waveforms on a per-channel basis and combined with the other pulses in an SMS slice group to form the final SMS-pTX pulse. Optimal spokes locations were derived from simulations.
RESULTS: Flip angle maps from presaturation TurboFLASH showed improvement of flip angle homogenization with 3-spoke pulses over CP-mode excitation (normalized root-mean-square error [NRMSE] 0.357) as well as comparable excitation homogeneity across the single-band (NRMSE 0.119), SMS-2 (NRMSE 0.137), and SMS-3 (NRMSE 0.132) 3-spoke pulses. The application of the 3-spoke SMS-3 pulses in a 48-slice GRE protocol, which has an in-plane resolution of 0.28 × 0.28 mm, resulted in a 50% reduction of scan duration (total acquisition time 6:52 min including reference scans).
CONCLUSION: Time-efficient flip angle homogenized high-resolution GRE imaging at 9.4T was accomplished by using slice-specific SMS-pTx spokes excitations. Magn Reson Med, 2016. © 2016 The Authors Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution NonCommercial License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.
Original language | English |
---|---|
Pages (from-to) | 1050–1058 |
Number of pages | 9 |
Journal | Magnetic Resonance in Medicine |
Volume | 78 |
Issue number | 3 |
DOIs | |
Publication status | Published - Sept 2017 |
Keywords
- ultrahigh-field MR
- parallel transmission
- magnitude least-squares
- spokes pulses
- simultaneous multislice
- high-resolution GRE SMS-pTX
- LEAST-SQUARES OPTIMIZATION
- TO-NOISE RATIO
- HIGH-FIELD MRI
- ABSORPTION RATE
- PULSE DESIGN
- RF PULSES
- HUMAN BRAIN
- PEAK POWER
- 7 TESLA
- EXPLICIT CONTROL