TY - JOUR
T1 - Extracorporeal Membrane Oxygenation Related Axillary Artery Cannulation and Aortic Hemodynamics
T2 - Insights From Computational Fluid Dynamics
AU - Coveliers, Jan
AU - Gasparotti, Emanuele
AU - Vignali, Emanuele
AU - Meani, Paolo
AU - Mazzoli, Marilena
AU - Kowalewski, Mariusz
AU - Huizinga, Eliza
AU - Gutta, Karthik
AU - Piccirillo, Giulia
AU - Huberts, Wouter
AU - Moradi, Hamed
AU - Di Mauro, Michele
AU - Holtackers, Robert J
AU - de Jong, Monique
AU - Gelsomino, Sandro
AU - Paparella, Domenico
AU - Körver, Erik
AU - Doddema, Arne
AU - Kawczinski, Michal
AU - Heuts, Sam
AU - Bidar, Elham
AU - Haxhiademi, Dorela
AU - Celi, Simona
AU - Lorusso, Roberto
N1 - Copyright © ASAIO 2026.
PY - 2026/2/2
Y1 - 2026/2/2
N2 - Axillary and subclavian artery cannulation for veno-arterial extracorporeal membrane oxygenation (V-A ECMO) offers clinical advantages but introduces complex flow dynamics within the aortic arch. This study employed patient-specific computational fluid dynamics (CFD) models derived from 10 computed tomography (CT)-based geometries to simulate ECMO flow via either the right axillary artery or left axillary artery access under varying levels of cardiac dysfunction. Three distinct flow behaviors were observed-descending-directed, arch-split, and retrograde aortic valve (AV)-directed patterns-depending on access site, support level, and aortic geometry. Right axillary artery access more frequently resulted in retrograde flow, particularly in type III arch configurations. In contrast, left axillary artery cannulation promoted antegrade or arch-split flow, independent of arch morphology. These findings suggest that left axillary artery access may provide more favorable antegrade flow hemodynamics in V-A ECMO, particularly when native cardiac output is impaired. Computational fluid dynamics modeling offers valuable insights to guide individualized cannulation strategies based on patient anatomy and cardiac function.
AB - Axillary and subclavian artery cannulation for veno-arterial extracorporeal membrane oxygenation (V-A ECMO) offers clinical advantages but introduces complex flow dynamics within the aortic arch. This study employed patient-specific computational fluid dynamics (CFD) models derived from 10 computed tomography (CT)-based geometries to simulate ECMO flow via either the right axillary artery or left axillary artery access under varying levels of cardiac dysfunction. Three distinct flow behaviors were observed-descending-directed, arch-split, and retrograde aortic valve (AV)-directed patterns-depending on access site, support level, and aortic geometry. Right axillary artery access more frequently resulted in retrograde flow, particularly in type III arch configurations. In contrast, left axillary artery cannulation promoted antegrade or arch-split flow, independent of arch morphology. These findings suggest that left axillary artery access may provide more favorable antegrade flow hemodynamics in V-A ECMO, particularly when native cardiac output is impaired. Computational fluid dynamics modeling offers valuable insights to guide individualized cannulation strategies based on patient anatomy and cardiac function.
U2 - 10.1097/MAT.0000000000002617
DO - 10.1097/MAT.0000000000002617
M3 - Article
C2 - 41627927
SN - 1058-2916
JO - ASAIO journal (American Society for Artificial Internal Organs : 1992)
JF - ASAIO journal (American Society for Artificial Internal Organs : 1992)
ER -