Dependence of treatment planning accuracy in peptide receptor radionuclide therapy on the sampling schedule

Christian Maass, Jan Philipp Sachs, Deni Hardiansyah, Felix M. Mottaghy, Peter Kletting, Gerhard Glatting*

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

Background: Peptide receptor radionuclide therapy (PRRT) plays an important role in the treatment of neuroendocrine tumors (NET). Pre-therapeutic dosimetry using the area under the measured time-activity curve (AUC) is important. The sampling schedule for this dosimetry determines the accuracy and reliability of the obtained AUC. The aim of this study was to investigate the effect of reduced number of measurement points (i.e., gamma camera image acquisition or serum measurements) on treatment planning accuracy in PRRT using In-111-labeled-diethylenetriaminopentaacetic acid-octreotide (DTPAOC; Octreoscan T). Methods: Pre-therapeutic biokinetic data of 15 NET patients were investigated using a recently developed physiologically based pharmacokinetic (PBPK) model. Two parameter sets were determined (standard or iterative approach) and used for calculation of time-integrated activity coefficients (TIACs) for the tumor, kidneys, liver, spleen, serum, and whole body. TIACs obtained using the full data sets were used as reference. To evaluate the effect of sampling on individual treatment planning, reduced sampling schedules were generated omitting either 1, 2, 3, or 4 organ and serum measurements or all serum measurements for each patient. Relative deviations (RDs) between these and reference TIACs were calculated and used as criterion for treatment planning accuracy. An RD <0.1 was considered acceptable. Results: When omitting serum measurements, TIAC accuracy remained acceptable (RD <0.1) for the standard approach. The kidney TIACs could be estimated for both approaches with acceptable RDs using two time points (t = 4 h; 2 d); tumor RDs were <0.3. The iterative approach reduced the range of RD, but did not further reduce the number of needed measurement points (i.e., to achieve an RD <0.1). For both approaches RDs for liver, spleen and whole body were larger than 0.1. However, in the clinical setting these RDs are less relevant as liver and spleen are not organs at risk due to the low absorbed doses. Conclusions: When using a priori information of a PBPK model structure combined with Bayesian information about PBPK model parameter distribution, the administered activity could be determined with acceptable accuracy using only two time points (4 h, 2 d) and thus allow a considerable reduction of needed data for individual dosimetry.
Original languageEnglish
Article number30
JournalEJNMMI Research
Volume6
DOIs
Publication statusPublished - 25 Mar 2016

Keywords

  • Physiologically based pharmacokinetic (PBPK) modeling
  • Peptide receptor radionuclide therapy (PRRT)
  • Optimal sampling schedule
  • Neuroendocrine tumor (NET)
  • Time-integrated activity coefficient (TIAC)

Cite this