Comparison of objective quality parameters between CTA and CTP angiographic reconstructions in ischemic stroke patients

M. M. Q. Robbe*, F. M. E. Pinckaers, A. H. H. Dirx, P. H. M. Voorter, W. H. van Zwam, B. A. J. M. Wagemans, A. A. Postma

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

Introduction: CT perfusion-angiographic reconstructions (CTP-AR) may be used for occlusion detection in ischemic stroke patients. Objective image quality of CTP-AR needs to be evaluated before implementation as it may affect occlusion detection. In this study, we aimed to assess the objective image quality, by means of contrast to noise ratio (CNR) and signal to noise ratio (SNR), of both CT-angiography and CT perfusion-angiographic reconstructions (CTP-AR). Methods: Patients with an ischemic stroke between September 2020 up to and including September 2021 who underwent both CT perfusion and CTA at baseline were included. CTP-AR was reconstructed from 1 mm CTP series at the peak arterial enhancement. Per patient, five ipsilateral and five contralateral regions of interest (ROI) were placed. Attenuation and standard deviation per ROI were used to calculate CNR and SNR. Differences in CNR and SNR between CTA and CTP-AR were tested using paired-sample t-tests. Results: In total, 195/239 patients were included. Both on the ipsilateral and contralateral side, the CNR was significantly higher on CTP-AR compared to CTA (P < .001 and P < .001, respectively). The SNR measured in the M1 was not significantly different between CTA and CTP-AR (ipsilateral: P = .68; contralateral: P = .63). The SNR, both on the ipsilateral and contralateral side, was significantly lower on CTP-AR compared to CTA in all parenchyma regions; the caudate nucleus (P < .001), lentiform nucleus (P < .001), centrum semiovale (P < .001), and the parenchyma adjacent to the M1 (P < .001). Conclusion: Image quality measures of CTP-derived angiographic reconstructions indicate higher CNR compared to CTA, but a lower SNR in non-angiographic structures.
Original languageEnglish
Article number100634
Number of pages6
JournalEuropean journal of radiology open
Volume14
DOIs
Publication statusPublished - 1 Jun 2025

Keywords

  • Ischemic stroke
  • Contrast-to-noise ratio
  • Signal-to-noise ratio
  • Occlusion detection
  • IMAGE QUALITY

Fingerprint

Dive into the research topics of 'Comparison of objective quality parameters between CTA and CTP angiographic reconstructions in ischemic stroke patients'. Together they form a unique fingerprint.