Comparison of central corneal thickness and anterior chamber depth measurements using three imaging technologies in normal eyes and after phakic intraocular lens implantation.

M. Doors, L.P. Cruysberg, T.T. Berendschot, J. de Brabander, F. Verbakel, C.A. Webers, R.M. Nuijts*

*Corresponding author for this work

    Research output: Contribution to journalArticleAcademicpeer-review

    Abstract

    BACKGROUND: The repeatability and interchangeability of imaging devices measuring central corneal thickness (CCT) and anterior chamber depth (ACD) are important in the assessment of patients considering refractive surgery. The purpose of this study was to investigate the agreement of CCT and ACD measurements using three imaging technologies in healthy eyes and in eyes after phakic intraocular lens implantation (pIOL). METHODS: In this comparative study, CCT and ACD were measured using anterior segment optical coherence tomography (AS-OCT), Orbscan II, and Pentacam in 33 healthy volunteers (66 eyes) and 22 patients (42 eyes) after pIOL implantation. Intraobserver repeatability was evaluated for all three devices in the healthy volunteer group. RESULTS: Pairwise comparison of CCT measurements showed significant differences between all devices (P < 0.001), except for the AS-OCT and Orbscan II in the healthy volunteer group (P = 0.422) and the Orbscan II and Pentacam in the pIOL group (P = 0.214). ACD measurements demonstrated significant differences between all pairwise comparisons in both groups (P </= 0.001). Intraobserver reliability was high for CCT and ACD measurements in the healthy volunteer group, with coefficients of variation ranging from 0.6% to 1.2% and 0.4% to 0.5% respectively. CONCLUSIONS: CCT and ACD measurements using AS-OCT, Orbscan II, and Pentacam demonstrated high intraobserver reliability. However, these devices should not be used interchangeably for measurements of CCT and ACD in healthy subject and patients after pIOL implantation.
    Original languageEnglish
    Pages (from-to)139-1146
    JournalGraefe's Archive for Clinical and Experimental Ophthalmology
    Volume247
    Issue number8
    DOIs
    Publication statusPublished - 1 Jan 2009

    Fingerprint

    Dive into the research topics of 'Comparison of central corneal thickness and anterior chamber depth measurements using three imaging technologies in normal eyes and after phakic intraocular lens implantation.'. Together they form a unique fingerprint.

    Cite this