Assessment of scattered and leakage radiation from ultra-portable X-ray systems in chest imaging: An independent study

  • Leonie E Paulis*
  • , Roald S Schnerr
  • , Jarred Halton
  • , Zhi Zhen Qin
  • , Arlene Chua
  • *Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

Ultraportable (UP) X-ray devices are ideal to use in community-based settings, particularly for chest X-ray (CXR) screening of tuberculosis (TB). Unfortunately, there is insufficient guidance on the radiation safety of these devices. This study aims to determine the radiation dose by UP X-ray devices to both the public and radiographers compared to international dose limits. Radiation dose measurements were performed with four UP X-ray devices that met international criteria, utilizing a clinically representative CXR set-up made with a thorax phantom. Scatter and leakage radiation dose were measured at various positions surrounding the phantom and X-ray tube, respectively. These measurements were used to calculate yearly radiation doses for different scenarios based on the median of all UP X-ray devices. From the yearly scatter doses, the minimum distances from the phantom needed to stay below the international public dose limit (1 mSv/year) were calculated. This distance was longest in the direction back towards the X-ray tube and shortest to the left/right sides of the phantom, e.g., 4.5 m and 2.5 m resp. when performing 50 exams/day, at 90 kV, 2.5 mAs and source skin distance (SSD) 1 m. Additional calculations including leakage radiation were conducted at a typical radiographer position (i.e., behind the X-ray tube), with a correction factor for wearing a lead apron. At 2 m behind the X-ray tube, a radiographer wearing a lead apron could perform 106 exams/day at 2.5 mAs and 29 exams/day at 10 mAs (90 kV, SSD 1 m), while keeping his/her radiation dose below the public dose limit (1 mSv/year) and well below the radiographer dose limit (20 mSv/year). In most CXR screening scenarios, the radiation dose of UP X-ray devices can be kept below 1 mSv/year by employing basic radiation safety rules on time, distance and shielding and using appropriate CXR exposure parameters.
Original languageEnglish
Article numbere0003986
JournalPLOS Global Public Health
Volume5
Issue number1
DOIs
Publication statusPublished - 24 Jan 2025

Fingerprint

Dive into the research topics of 'Assessment of scattered and leakage radiation from ultra-portable X-ray systems in chest imaging: An independent study'. Together they form a unique fingerprint.

Cite this