Accuracy of the Precision Saw versus the Sagittal Saw during total knee arthroplasty: A randomised clinical trial

Peter Z. Feczko*, Hugo J. P. Fokkenrood, Tijmen van Assen, Patrick Deckers, Pieter J. Emans, Jacobus J. Arts

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

Background: The aim of this study was to compare the accuracy of the oscillating tip saw system (Precision Saw = PS) with the more conventional fully oscillating blade system (Sagittal Saw = SS) during computer-assisted total knee arthroplasty (CAS-TKA).

Methods: A prospective, randomised, controlled trial included 58 consecutive patients who underwent primary CAS-TKA and were randomly assigned in the PS group or the SS group to compare the accuracy of both blades. The primary outcome was the difference between the intended cutting planes and the actual cutting planes in degrees () in two planes of both the femur and the tibia. The secondary outcome was total surgery time.

Results: Tibia: In the W-plane no significant differences were registered for the mean absolute deviation (p = 0.28). The PS was more accurate in the AP-plane (p = 0.03). Femur: The PS showed significantly fewer mean absolute deviations in the W-plane (p = 0.03); however, the SS revealed better accuracy in the FE-plane (p = 0.04). The difference in the surgery time between the groups was not statistically significant (p = 0.45). Two outliers were measured using the SS, while seven outliers were detected using the PS.

Conclusion: The Precision Saw is not proven to be overall more accurate than the Sagittal Saw. Significantly better accuracy was shown with the PS in the two cutting planes, with the exception of one cutting plane that favoured the SS. Greater number of outliers were found using the PS. (C) 2017 Elsevier B.V. All rights reserved.

Original languageEnglish
Pages (from-to)1213-1220
Number of pages8
JournalKnee
Volume24
Issue number5
DOIs
Publication statusPublished - Oct 2017

Keywords

  • Orthopaedic saw
  • Total knee arthroplasty
  • Navigation
  • Accuracy
  • Surgery time
  • CUTTING ERRORS
  • NAVIGATION SYSTEM
  • DISTAL FEMUR
  • REPLACEMENT
  • ALIGNMENT

Cite this