A standardized evaluation of artefacts from metallic compounds during fast MR imaging

  • Shumei Murakami
  • , Rinus G. Verdonschot*
  • , Miyoshi Kataoka
  • , Naoya Kakimoto
  • , Hiroaki Shimamoto
  • , Sven Kreiborg
  • *Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

Objectives: Metallic compounds present in the oral and maxillofacial regions (OMRs) cause large artefacts during MR scanning. We quantitatively assessed these artefacts embedded within a phantom according to standards set by the American Society for Testing and Materials (ASTM).textlessbrtextgreater Methods: Seven metallic dental materials (each of which was a 10-mm(3) cube embedded within a phantom) were scanned [i.e. aluminium (Al), silver alloy (Ag), type IV gold alloy (Au), gold-palladium-silver alloy (Au-Pd-Ag), titanium (Ti), nickel-chromium alloy (NC) and cobalt-chromium alloy (CC)] and compared with a reference image. Sequences included gradient echo (GRE), fast spin echo (FSE), gradient recalled acquisition in steady state (GRASS), a spoiled GRASS (SPGR), a fast SPGR (FSPGR), fast imaging employing steady state (FIESTA) and echo planar imaging (EPI; axial/sagittal planes). Artefact areas were determined according to the ASTM-F2119 standard, and artefact volumes were assessed using OsiriX MD software (Pixmeo, Geneva, Switzerland).textlessbrtextgreater Results: Tukey-Kramer post hoc tests were used for statistical comparisons. For most materials, scanning sequences eliciting artefact volumes in the following (ascending) order FSE-T-1/FSE-T-2 textless FSPGR/SPGR textless GRASS/GRE textless FIESTA textless EPI. For all scanning sequences, artefact volumes containing Au, Al, Ag and Au-Pd-Ag were significantly smaller than other materials (in which artefact volume size increased, respectively, from Ti textless NC textless CC). The artefact-specific shape (elicited by the cubic sample) depended on the scanning plane (i.e. a circular pattern for the axial plane and a "clover-like" pattern for the sagittal plane).textlessbrtextgreater Conclusions: The availability of standardized information on artefact size and configuration during MRI will enhance diagnosis when faced with metallic compounds in the OMR.
Original languageEnglish
JournalDentomaxillofacial Radiology
Volume45
Issue number8
DOIs
Publication statusPublished - 2016
Externally publishedYes

Fingerprint

Dive into the research topics of 'A standardized evaluation of artefacts from metallic compounds during fast MR imaging'. Together they form a unique fingerprint.

Cite this